FINAL TOOLBOX
04 // Interactive Formula Compendium // CHEMISTRY-THERMOA = ε · l · c = −log₁₀(I / I₀)

Beer-Lambert Optical Absorbance Law

The Beer-Lambert law A = εlc relates dimensionless absorbance A = −log₁₀(I/I₀) linearly to molar extinction coefficient ε (L·mol⁻¹·cm⁻¹), optical path length l (cm), and solute concentration c (mol/L).

Formula Reference
Main Formula
A = ε · l · c = −log₁₀(I / I₀)
Rearranged Forms
T = I / I₀ = 10^(−A)
c = A / (ε · l)
Variables & Units
Variables & Units — Beer-Lambert Optical Absorbance Law
SymbolNameUnit
AOptical Absorbance AAU
epsMolar Absorptivity εL/(mol·cm)
lCuvette Path Length lcm
cMolar Concentration cmol/L
Inputs & Parameters
[L/(mol·cm)]
[cm]
[mol/L]
Calculated Result (Optical Absorbance A)
A =0.622AU

Formula Notes & Background

  • Differential attenuation dI = −σ N I dx integrated across thickness l yields I = I₀ e^(−σNl), converted to base-10 absorbance A = εlc.
  • Spectrophotometers are most accurate for A between 0.1 and 1.5 AU (transmittance 79% down to 3.2%).

Step-by-Step Examples

#1: NADH solution at c = 1.0 × 10⁻⁴ M (100 μM) measured at 340 nm (ε = 6,220 M⁻¹cm⁻¹) in a standard 1.0 cm quartz cuvette.
ε = 6,220 L/(mol·cm)l = 1.0 cmc = 1.0×10⁻⁴ M
A = 6,220 × 1.0 × 0.0001
→ A = 0.622 AU (Transmittance T = 10^(−0.622) = 23.9%)
#2: DNA concentration check at 260 nm where A = 1.00 corresponds to 50 μg/mL dsDNA in a 1 cm path.
Measured A₂₆₀ = 0.48 AUl = 1.0 cm
Concentration = 0.48 × 50 μg/mL
→ 24.0 μg/mL dsDNA
Was this tool helpful?

Frequently Asked Questions

Why does Beer-Lambert linearity deviate above A > 2.0?+

At A = 2.0 only 1% of light transmits (0.1% at A = 3.0), so stray instrument light, refractive index shifts, and solute dimerization cause apparent flattening.

What does A = 1.0 mean in terms of transmitted light?+

A = 1.0 means 90% of incident photons are absorbed and exactly 10% are transmitted (I/I₀ = 0.10).

Why are UV measurements below 300 nm run in quartz rather than plastic/glass cuvettes?+

Standard borosilicate glass and polystyrene strongly absorb UV wavelengths below 320 nm, whereas fused quartz is transparent down to 190 nm.